The Estriol Competitive ELISA quantitates estriol in serum, plasma, urine, saliva, dried fecal extracts or or cell culture medium.∣Principle of the method: The Estriol Competitive ELISA research-use-only kit is designed to quantitatively measure…
The Estriol Competitive ELISA quantitates estriol in serum, plasma, urine, saliva, dried fecal extracts or or cell culture medium.∣Principle of the method: The Estriol Competitive ELISA research-use-only kit is designed to quantitatively measure estriol independent of species. An estriol standard is provided to generate a standard curve for the assay and all samples are read off a user-generated standard curve. Standards or diluted samples are pipetted into a clear microtiter plate coated with anti-rabbit IgG antibodies. Anti-rabbit estriol antibodies and estriol-peroxidase conjugate is added to the wells. The estriol-peroxidase conjugate and any estriol in the sample will compete to bind to the anti-rabbit estriol antibodies. After incubation, the plate is washed and substrate is added. The substrate reacts with the bound estriol-peroxidase conjugate. After a 30 min incubation, the reaction is stopped and the intensity of the generated color is detected in a microtiter plate reader at 450 nm. The intensity of the generated color corresponds inversely to the amount of estriol present in the sample.∣Rigorous validation: Each manufactured lot of this ELISA kit is quality tested for criteria such as sensitivity, specificity, precision, and lot-to-lot consistency. See manual for more information on validation.
L-estriol is the primary estrogen in the blood and urine of pregnant females.Serum estrolic determinations reflect the state of the fetus and are used to assess fetal distress and placental function.
Sensitivity
Sensitivity of the dosage: 5.33 pg/ml.
Measurement range
Dose range: 19.66 to 12.000 pg/ml.
Duration of dosing
Total duration of dosing: 2 h 30 min.
Reproducibility
Coefficient of variation intra-dose 10%, inter-dose 12%.
Technical specifications
| General |
| Brand | Invitrogen |
| Supplier ref. (Fisher Scientific) | 17164853 |
| Specifications |
| Accession number | C18H24O3; Estratriol; Estriol; Oestriol; Ovestin; Ovestrion |
| Dose range | 19.66 to 12.000 pg/ml |
| Sensitivity of the dosage | 5.33 pg/ml |
| Conjugate | HRP |
| Product type | ELISA |
| For use with (equipment) | Colorimetric microplate reader |
| Coefficient of inter-dose variation | 12% |
| Coefficient of intra-dose variation | 10% |
| Contents of kits | 96 well plate coated with anti-IgG Standard Dose buffer concentrate Capture anticorps HRP-conjugate antigen Wash pad concentrate Substrate TMB Discontinue solution Thermosceller for plates |
| Marker or dye | HRP |
| Quantity | 96 dosages |
| Regulatory status | Use exclusively for research |
| Target | Oestriol |
| Test time | 45 min |
| Physical characteristics |
| Volume of sample | Serum, 50 µl; plasma, 50 µl; fecal extract, 50 µl; urine, 12.5 µl; supernatant, 50 µl |
| Storage and shipping |
| Storage conditions | -20°C |
| Total duration of dosing | 2 h 30 min |
Origin and customs
| Manufacturer ref. | EIAEST |
|---|
| Country of origin | United States |
|---|
| HS customs code | 3002 12 00 00 |
|---|
A question or a specific configuration?