Laboratory multiparameter Hanna edge® HI2030-02 – conductivity kit

Ref.: BL-EQ-134693Manufacturer ref.: HI2030-02
€880.00 excl. VAT

Ultra-fine edge® instrument delivered in conductivity kit: 4-ring platinum probe, EC, TDS and salinity over a very wide range, memory of 1000 recordings.

Accuracy (salinity)
+/-1 % des mesures
Accuracy (TDS)
+/-1 % des mesures
Calibration modes
TDS, salinité, EC
Range (salinity)
0.0 to 400.0 % de NaCl ; 0.01 to 42.00 PSU ; 0.0 to 80.0 g/l
Range (TDS)
0.00 to 14.99 mg/l (ppm) ; 15.0 to 149.9 mg/l (ppm) ; 150 to 1499 mg/l (ppm) ; 1.50 to 14.99 g/l ; 15.0
Brand
Hanna Instruments

EC Dedge® kit: from pure water to brine with a single probe.

Le Hanna edge® HI2030 is the version conductivity edge® multiparameter: a 13 mm thick instrument with a 5.5-inch screen, which can be used by hand, on a mat or on a wall. It measures conductivity, dissolved solids (TDS), salinity and temperature.

Sa HI763100 digital probe with 4 platinum rings covers a very wide range and resists polarization. The basic mode simplifies routine measurements, the standard mode opens all settings: temperature coefficient, TDS factor, measurement without compensation for absolute conductivity.

4 rings probe

HI763100 platinum probe: wide range and lack of polarization.

Tablet format

13 mm thick and 250 g: by hand, on a mat or on a wall.

1000 records

BPL data, date and time; export to USB drive or PC.

Features

  • Conductivity from 0.00 µS/cm to 200.0 mS/cm (up to 500.0 mS/cm in absolute)
  • TDS and salinity (% NaCl, PSU, g/L)
  • EC calibration on 6 memorized and offset standards at 0.00 µS/cm
  • Adjustable temperature coefficient from 0.00 to 6.00 %/°C, option without compensation
  • Sensitive keypad, not sensitive to projections
  • Basic (simplified display) or standard (full settings) mode
  • Memory of 1000 records with BPL data, date and time
  • Export on USB drive and PC connection by micro-USB
  • Paillass docking station with electrode holder and wall support provided
  • Evolutive: compatible pH electrodes and dissolved oxygen probes

Applications

Purified waterQuality controlFoodProcess waterSeawaterResearch laboratories

Technical specifications

Conductivity
Range0.00 to 29.9 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.9 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm in absolute conductivity
Resolution0.01 µS/cm ; 0.1 µS/cm ; 1 µS/cm ; 0.01 mS/cm ; 0.1 mS/cm
Accuracy±1 % of reading (±0.05 µS/cm or 1 digit)
CalibrationCell constant over 6 standards (84 µS/cm, 1413 µS/cm, 5.00 – 12.88 – 80.0 – 111.8 mS/cm); offset at 0.00 µS/cm
Temperature compensationAutomatic from -5 to 100 °C or disabled; coefficient 0.00 to 6.00 %/°C (1.90 %/°C by default)
TDS
Range0.00 to 14.99 mg/L up to 100.0 g/L; 400.0 g/L in absolute TDS (factor 0.80)
Accuracy±1 % of reading (±0.03 mg/L or 1 digit)
Conversion factor0.40 to 0.80 (default 0.50)
Salinity
% NaCl0.0 to 400.0 per cent, ±1 per cent accuracy of reading
Practical salinity2.00 to 42.00 PSU
Seawater0.0 to 80.0 g/L
Temperature
Range-20.0 to 120.0 °C
Resolution0.1 °C
Accuracy±0.5 °C
Instrument
Probe providedHI763100, 4 platinum rings, integrated temperature sensor
Screen5.5 inches LCD
MemoryUp to 1000 records (200 on demand, 200 on stability, 600 at intervals in 100 lots)
Connectivity1 micro-USB port (PC and recharge); 1 USB port (storage key)
Sensor inputJack 3.5 mm, digital probes
Battery / battery autonomyIntegrated rechargeable battery, up to 8 hours continuous use
Power supply5 V DC adapter (provided)
Terms of use0 to 50 °C, 95 % max. HR without condensation
Dimensions202 × 140 × 12.7 mm
Weight250 g

Origin and customs

Manufacturer ref.HI2030-02
Country of originRomania
HS customs code9027 89 30 00

Scope of delivery

  • Instrument edge®
  • Conductivity probe HI763100
  • Bags of calibration solutions 1413 µS/cm and 12 880 µS/cm
  • Paillass docking station with electrode holder and wall support
  • USB cable and 5 V adapter
  • Instructions for use and quality certificate

Accessoires compatibles

  • HI763100 — Sonde de conductivité numérique 4 anneaux
  • HI7037L — Solution d’étalonnage de salinité (eau de mer 100 % NaCl)
  • HI11310 — Électrode pH numérique pour edge®
  • HI764080 — Sonde d’oxygène dissous numérique pour edge®
A question or a specific configuration?
Contact us for a quote.