Traffic cooler IKA RC 5 control – 1,400 W, up to −30 °C, wireless controller

Ref.: BL-EQ-035535
Price on request
Custom pricing depending on configuration and quantity.

High-power circulation cooler (1,400 W at 20 °C) for external applications up to −30 °C, with natural fluid R290, with regulated PEEK pump and 7 l insulated bath, wireless controller.

Supplier ref.: 16238181

Ambient temperature
5 to 32°C
Flow
31 l/min
Pump type
Pompe de pression/aspiration
Temperature range
−30 to 80°C
Max.
0.61 bars
Brand
IKA

1,400 W of natural cold, programmable ramps and wireless steering.

Le IKA RC 5 control is a powerful circulation cooler for external cooling applications up to and including −30 °C. It uses natural refrigerant R290 and offers a high power density for a reduced footprint on the mat.

Its regulated speed PEEK pump continuously adjusts pressure and flow. The 7 l insulated bath has an integrated funnel and a drain valve that can be connected to a hose to empty the bath without contact with the fluid. The WiCo Wireless Controller allows you to program 10 10-step temperature ramps, with timer, set-off signal and degassing function.

High refrigerating capacity

Regulated speed refrigeration system for rapid cooling of large volumes.

Natural fluid R290

An environmentally friendly refrigerant with high power density.

WiCo Controller

Wireless control, 10 programmable ramps and external PT 100 probe.

Features

  • Refrigeration system at regulated speed, 1,400 W at 20 °C
  • Natural refrigerant R290
  • Working range from -30 °C to ambient, up to 80 °C with external heating
  • PEEK pump at regulated speed, 0.6 bar and 31 l/min
  • 7 l insulated bath with built-in funnel and drain valve
  • Digital level indicator
  • Frontal clip for cleaning the air filter
  • RS 232 and USB interfaces
  • WiCo Wireless Controller: 10 10-step ramps, timer, confirmation mode, degassing
  • LEMO socket for external PT 100 probe, multiple inputs/outputs (electrovan, alarm, watch)

Applications

Reactor coolingLarge-scale rotary evaporationCondensersBioreactorsSynthetic chemistryPilot processes

Technical specifications

Performance
Refrigeration capacity1 400 W à 20 °C ; 950 W à 0 °C ; 450 W à −20 °C ; 200 W à −30 °C
Working range-30 °C at ambient temperature (up to 80 °C with external heating)
Stability (DIN 12876)±0.1 K
Pump: max. pressure / suction.0.61 bar / 0.45 bar
Max.31 l/min
Filling volume5.0 to 7.0 l
Refrigerant fluidR290 (90 g)
Operation
DisplayTFT, resolution 0.01 K, WiCo wireless controller
ControlPID, probe PT 100
InterfacesRS 232, USB
Safety
Protection index (DIN EN 60529)IP 21
Sound level61 dB(A)
Dimensions and weight
Dimensions (l × H × P)310 × 546 × 490 mm
Weight37.5 kg
Power supply
Voltage/frequency230 V, 50/60 Hz
Power absorbed1 100 W

Scope of delivery

  • Traffic cooler RC 5 control
  • WiCo Wireless Controller
  • Temperature sensor PT 100

Accessoires compatibles

  • PT100.30 — Sonde de température PT 100 à connecteur LEMO
A question or a specific configuration?
Contact us for a quote.

Supplier ref.: 16238181