IP67 portable waterproof conductivity: conductivity from 0.001 µS/cm to 1000 mS/cm, TDS, resistivity, salinity and temperature, USP mode <645>, 4-ring platinum probe.
Conductivity
0,001 µS/cm à 1000 mS/cm
Accuracy
±1 % de la lecture
Paramètres
EC, TDS, résistivité, salinité, °C
Compliance
USP <645>, BPL
Protection rating
IP67
Brand
Hanna Instruments
Conductivity, TDS, resistivity and salinity in a single field device, compliant with USP pharmacopoeia <645>.
Le HI98192 Hanna Instruments is a professional portable conductimeter for routine measurements and expertise in the field, in production or in the laboratory. Its wide range, from 0.001 µS/cm to 1000 mS/cm, also allows the calculation of TDS, resistivity and three salinity values.
Its mode USP <645> makes it suitable for the control of purified water and water for injections. Waterproof, compact and lightweight, it comes complete in case with its 4 platinum rings probe.
Multiparameter
Conductivity, resistivity, TDS, salinity and temperature with one probe.
Compliant USP <645>
Dedicated mode for purified water and water for injections.
Calibration 5 points
7 memorized standards from 84 µS/cm to 111.8 mS/cm for high accuracy.
10 measurement profiles
Reference temperature, compensation and TDS factor stored by application.
Memorandum 1,400 measurements
400 on demand and 1,000 in automatic recording, export by micro-USB.
IP67 Water Resistance
Robust and handy case for field measurements.
Features
Conductivity, TDS, resistivity, salinity (% NaCl, PSU, sea water) and temperature
Automatic calibration up to 5 points
Linear or non-linear temperature compensation (ISO/DIN 7888)
Adjustable cell constant
Automatic or manual range selection
10 custom measurement profiles
Automatic recording (5 seconds to 180 minutes)
Opto-isolated microUSB port, HI92000 software
Applications
Purified water and pharmaceuticals (USP <645>)Water and environment analysisQuality control in productionAnalysis laboratoriesAquaculture and sea waterField measures
Technical specifications
Conductivity
Range
0.001 µS/cm at 1000 mS/cm (absolute display beyond 400 mS/cm)
Resolution
0.001 µS/cm to 0.1 mS/cm depending on the range
Accuracy
±1 % of reading (±0.01 µS/cm or 1 digit)
Calibration
Up to 5 points, 7 memorized standards
TDS
Range/accuracy
0.00 mg/L to 400.0 g/L / ±1 % of reading
Factor
0.40 to 1.00
Resistivity
Range
1.0 ι cm to 100.0 Mιcm
Salinity
Ranges
0.0 to 400.0 % NaCl; 0.00 to 42.00 PSU; 0.00-80.00 g/L
Temperature
Range/accuracy
-20.0 at 120.0 °C / ±0.2 °C (excluding probe)
Compensation
None, linear (0 to 10 %/°C) or non-linear ISO/DIN 7888; ref. 15, 20 or 25 °C
Instrument
Probe provided
HI763133, 4 platinum rings, DIN, 1.5 m cable, max. 70 °C.
Memory
400 on demand, 1,000 automatic, 10 profiles
Connectivity
Micro-USB, Opto-isolated software HI92000
Power supply / Autonomy
4 AA 1.5 V batteries / approx. 100 hours (25 hours with backlight)
Protection
IP67
Dimensions/weight
185 × 93 × 35.2 mm / 400 g
Origin and customs
Manufacturer ref.
HI98192
Country of origin
Romania
HS customs code
9027 89 30 00
Scope of delivery
HI98192 conductimeter
EC/°C probe HI763133 (4 platinum rings, temperature sensor, DIN connector, 1.5 m cable)
Calibration solutions 1413 µS/cm and 111.8 mS/cm (230 mL each)