| General |
| Brand | Thermo Scientific |
| Description | Orion Star T930 Electrode-Free Ion Titrator |
| Model | Orion Star T930 |
| Range (specific ions) | 0.0001 to 19990 |
| Range (mV) | -2000.0 to +2000.0 mV |
| Relative humidity range | 5 to 85 %, non-condensing |
| Performance |
| Clarification (specific ions) | ± 0.2 mV or ± 0.05 per cent of the measurement, whichever is greater |
| Accuracy (temperature) | ±0.2°C |
| Accuracy (mV) | ±0.2 mV |
| Advanced ionometric functions | Multiple known addition (MKA); automated calibration and calculation of sample concentration by adding standard aliquots to the sample using the burette |
| Alarm outputs | Notifications for the end of a titration cycle, maximum titrating volume, complete data log, calibration alarm, maintenance recall |
| Backlight option | Yes, adjustable |
| White Options | Fixed or variable value by means of titration |
| Functions relating to the burette | Automatic rinse cycles for rinsing and separate distribution volume with continuous option; exceeds |
| Sizes of burettes | 10 ml, 20 ml (included) or 50 ml |
| Channels | 1 |
| Titration cycles | 1 to 5 cycles, possibility of excluding results from the cycle of mean results and ETR calculations |
| Details of data management | Export of data log in CSV or PDF format, short or long format |
| Data points | 100 data sets for each: sample titration, titrant standardization, calibration and direct measurement |
| Equivalency points | 1 to 2 |
| For use with (application) | Ion concentration titrations, including salt in food, calcium in juice and milk, total hardness of drinking water and wastewater, chloride in drinking water and wastewater, ammonia and total nitrogen of Kjeldahl (TKN) in wastewater and surfactants in consumer products. |
| Available functions | Sample titration, titrant standardization, calibration and direct measurement |
| IP Index | IP 51 |
| Includes | 20 ml burette, agitation probe, distribution probe, standard tubing kit, plastic bottle 1 L, stopper with two ports for bottle GL38 and drying tube, computer cable, USB drive documentation, 110-240 V power supply |
| Entry | BNC (selective electrode ions), spindle (reference electrode), 8 pin MiniDIN (temperature probe), actuator input (agitation probe) |
| Languages | Chinese, English, French, German, Italian, Japanese, Korean, Portuguese and Spanish |
| Memory | Non-volatile memory keeps data logs and settings |
| Methods | Up to 10 methods with optional password protection; import/export to methods via USB flash drive |
| Outputs | 3 USB ports for printer, USB flash drive and computer software connections |
| Accuracy | 0.5 % ETR, depending on the environment and handling conditions |
| Predefined turning points | 1 to 3 |
| Type of probe | Selective ion electrode (ISE), reference half-cell electrode, ATC temperature probe, agitator probe |
| Resolution | The burette engine offers advanced micro-step technology that delivers 25.600 micro-steps per engine turn for easy and accurate positioning of the burette (more than 2 million micro-steps throughout the burette race) |
| Resolution (specificions) | Minimum 0.0001, 1 to 4 significant digits (selected by the user) |
| Resolution (temperature) | 0.1°C |
| Resolution (mV) | 0.1 mV |
| Sample ID | Automatic, manual or deactivated increase |
| Speeds of the agitator probe | 5 speeds selectable per user (250 to 3700 rpm) |
| Temperature (metric) in operation | 5 to 40°C |
| Temperature range (metric) | -5.0 to 100.0°C ; 23.0 to 212°F |
| Temperature selection | Manual or automatic |
| Time and Date | Yes, with the non-volatile backup battery |
| Determination of the titrated solution | Standard titration or manual addition of concentration |
| Control of the titration process | Routine, fast, cared for or defined by the user |
| Titration Configuration Assistant | Yes, using the workflow Start a new titration |
| Titration technique | Pre-determined point of equivalence or end point |
| Types of titration | Direct or return titration |
| Software that can be updated | Yes, thanks to a USB drive |
| Display type | 5.7 inches capacitive colour touch screen, 640 x 480 resolution |
| Type | Ion titrator |
| Physical characteristics |
| Dimensions (L x W x H) | 10” x 16” x 14” ; 25.4 cm x 40.6 cm x 35.6 cm |
| Material | ABS plastic |
| Electrical |
| Power supply | 100/240 V, 50/60 Hz |
| Compliance and warranty |
| Calibration | Ion concentration (ISE) and temperature |
| Calibration points (ionometry) | 1 to 5 points |
| Certifications/Compliance | EC, TUV 3-in-1, FCC, EN/IEC61326-1, IEC 61010 |
| Calibration of the temperature sensor | Calibration of 1 optional offset point |
| Warranty | 1 year warranty (replacement) |