Add conductivity, TDS, salinity or resistivity measurement capabilities to your Thermo Scientific™ Orion™ VERSA STAR™ measuring instrument.
Brand
Thermo Scientific
Reference
VSTAR-CND
Brand
Thermo Scientific
Add conductivity, TDS, salinity or resistivity measurement capabilities to your Thermo Scientific™ Orion™ VERSA STAR™ measuring instrument.
Add conductivity, TDS, salinity or resistivity measurement capabilities to your measuring instrument with the Thermo Scientific™ Orion™ VERSA STAR™ conductivity module. This module offers a choice of practical salinity curves or natural sea water for salinity measurements and selectable reference measurement temperatures of 5°, 10°, 15°, 20° or 25°C with linear, nonlinear, nLFu or EP curve options for conductivity measurements.
Specifications
Simple language on-screen invitations and versatile quick keys to facilitate selection
Interface and data
Interface in several languages supporting English, Spanish, French, Italian, German and Chinese.
Non-volatile memory that can store up to 2...
Non-volatile memory that can store up to 2,000 data points with an indication of the time and date.
Two USB ports
Two USB and RS232 ports to facilitate data transfer and software update of the measuring instrument.
Data analysis software
Data analysis software for transferring stored results or displaying measurements in real time.
Samples and capacity
Direct control of two Orion Star™ (Ref. 096019, sold separately) for mixing samples without using agitation plate.
Features
Simple language on-screen invitations and versatile quick keys to facilitate selection
Interface in several languages supporting English, Spanish, French, Italian, German and Chinese
Non-volatile memory capable of storing up to 2,000 data points with time and date indication
Two USB and RS232 ports to facilitate data transfer and software update of the measuring instrument
Data analysis software for transferring stored results or displaying measurements in real time
Direct control of two Orion Star™ (Ref. 096019, sold separately) for mixing samples without using agitation plate
Electrode arm and probe support of new design for holding and placing probes in samples
Universal sector adapter with virtually all AC power sources
IP54 class enclosure supporting splashes
3-year warranty (replacement of the measuring instrument)
Technical specifications
General
Brand
Thermo Scientific
Description
Conductivity module / TDS / resistivity / salinity / temperature Orion VERSA STAR
Model
Conductivity module / TDS / resistivity / salinity / temperature Orion Versa Star
Range (conductivity)
0.001 µS/cm at 3.000 mS/cm
Range (resistivity)
2 ohms to 100 meg-ohms
Range (salinity)
0.01 to 80.0 ppt equivalent NaCl; 0.01 to 42 ppt practical salinity
Range (TDS)
0 to 200 ppt
Relative humidity range
5 to 85 % non-condensing
Range of TDS factors
Linear 0.01 to 10.00
Temperature range (metric)
-5 to +105°C
Reference temperature (metric)
5, 10, 15, 20, 25°C
Performance
Precision (conductivity)
0.5 per cent of reading ±1 digit > 3 μS; 0.5 per cent of reading ±0.01 μS ≤ 3 μS
Accuracy (resistivity)
0.5 per cent of measurement ± 1 digit
Accuracy (salinity)
± 0.1 ppt
Accuracy (TDS)
0.5 per cent of measurement ± 1 digit
Accuracy (temperature)
±0.1°C
Alarm outputs
Set-up point alarm, high/low limit alarm, calibration alarm
Cell Constant
0.001 to 199.9
Compatible probes
Two or four cell conductivity with integrated temperature
IP Index
IP-54
Includes
Versa Star conductivity module and printed Quick Start Guide
Entry
8 pin Mini-DIN (conductivity with temperature sensor)
Measurement modes
Conductivity, TDS, salinity or resistivity with temperature
Methods
10 per channel with password protection
Type of probe
Conductivity probe to two or four cells - to be purchased separately
Resolution (conductivity)
0.001 µS minimum, 4 significant figures at least
Resolution (resistivity)
2 ohm-cm
Resolution (salinity)
0.01 ppt
Resolution (TDS)
4 significant figures
Resolution (temperature)
0.1°C
Methods of measuring salinity
Practical salt or natural sea water (NaCl equivalent)
Stability
Quick, medium, slow or Smart Stability parameter selectable; Smart Averaging parameter disabled or automatically selectable
Thermal compensation
Linear, nLFn (non-linear ultra-pure water), nLFu (non-linear ultra-pure water), EP (discharge with warning) or shutdown
Temperature (metric) in operation
5°C to 45°C (ambient temperature)
Temperature selection
Manual or automatic with temperature sensor ATC
Type
Conductivity measurement module
Physical characteristics
Dimensions (L x W x H)
12°cm x 3°cm x 9,5°cm (4,7°in x 1,2°in x 3,8°in.)
Weight (metric)
0.2 kg
Compliance and warranty
Calibration
Conductivity with calibration and temperature editing option
Calibration points
Up to 6
Calibration of the temperature sensor
Temperature calibration in 1 point
Warranty
Replacement of the measuring instrument for three years
Origin and customs
Manufacturer ref.
VSTAR-CND
Country of origin
Indonésie
HS customs code
9027 90 00 00
A question or a specific configuration? Contact us for a quote.