EasyStep Human sST2(Soluble Suppression Of Tumorigenicity 2) ELISA Kit – 5 x 96 tests

Ref.: BL-RE-950363Manufacturer ref.: ELK008ES-5x96T 5 × 96 tests
€2,659.00 excl. VAT

EasyStep Human sST2(Soluble Suppression Of Tumorigenicity 2) ELISA Kit, 5 x 96 tests. For research use only.

Format
5 × 96 tests (5 plates of 96 wells)
Species
Human
Target
sST2 (Soluble Suppression Of Tumorigenicity 2)
Type of dosage
ELISA sandwich
Sensitivity
0.25 ng/mL
Brand
ELK Biotechnology

Quantitative ELISA human sST2 sandwich (serum, plasma), result in 1.5 h.

This ELISA EasyStep ELK Biotechnology kit allows the quantitative in vitro determination of human sST2 (Solvable Removal Of Tumorigenicity 2) in the following samples: serum and plasma. The EasyStep range reduces the protocol to one step of incubation.

The assay is based on the ELISA sandwich technique: the microtitration plate is pre-coatted with a specific sST2 antibody. Standards or samples are added, followed by a detection antibody conjugated with peroxidase (HRP), in one step. After addition of the TMB substrate and cessation of the sulfuric acid reaction, the optical density is measured at 450 nm ± 10 nm; the concentration is determined using the standard curve.

Research-only product (RUO), not intended for diagnosis.

Sensitivity

Sensitivity of 0.25 ng/mL.

Range of dosage

Measurement range from 0.47 to 30 ng/mL (highest standard: 30 ng/mL).

Samples

Serum and plasma.

Rapidity

Result in 1.5 h, protocol in one step.

Specificity

Sandwich format, human sST2 specific antibodies.

Reproducibility

Intra-test accuracy CV < 8%, inter-test CV < 10%.

Technical specifications

General
BrandELK Biotechnology
RangeEasyStep
Specifications
Format5 × 96 tests (5 plates of 96 wells)
SpeciesHuman
TargetsST2 (Soluble Suppression Of Tumorigenicity 2)
Type of dosageELISA sandwich
Performance
Sensitivity0.25 ng/mL
Range of dosage0.47 – 30 ng/mL
Standard (maximum concentration)30 ng/mL
Duration of dosing1.5 h
Readability wave length450 nm ± 10 nm
Specifications
SamplesSerum and plasma
Research areaCD and accession molecules
Alternative namess-ST2
Storage and shipping
ShippingWith ice loaves, without carboglace
Compliance
Useresearch use only (RUO)
Validation
Intra-test accuracyCV < 8 %
Inter-test accuracyCV < 10 %
Standard curve (example)30.00 ng/mL : DO 2.035; 15.00 ng/mL : DO 1.502; 7.50 ng/mL : DO 1.115; 3.75 ng/mL : DO 0.892; 1.88 ng/mL : DO 0.563; 0.94 ng/mL : DO 0.239; 0.47 ng/mL : DO 0.213; 0.00 ng/mL : DO 0.096
Recovery (doped samples)serum (n=5): 95-108% (average 101%); EDTA plasma (n=5): 80-95% (average 87%); heparin plasma (n=5): 82-97% (average 89%)
Linearity (dilutions)serum (n=5): 1:2 90-99 %, 1:4 83-97%, 1:8 81-103%, 1:16 89-101%; EDTA plasma (n=5): 1:2 88-99 %, 1:4 81-93%, 1:8 89-97%, 1:16 85-94%; heparin plasma (n=5): 1:2 87-96 %, 1:4 82-95 %, 1:8 93-102%, 1:16 87-96 %
Conservation
Storage of kitKit not open: 2-8 °C for 6 months, or -20 °C for 1 year; avoid freeze-thaw cycles.
Safety data sheetDownloadable with a B-LABO account

Origin and customs

Manufacturer ref.ELK008ES-5x96T
Country of originChina
HS customs code3822 19 00 00

Scope of delivery

  • 5 kits of 96 tests, each containing:
  • Pre-coated microplate: 12 strips × 8 wells (conservation 4 °C or -20 °C)
  • Microplate of standards: 2 strips × 8 wells (conservation 4 °C or -20 °C)
  • HRP conjugate antibody (100×): 70 μL (storage 4 °C or -20 °C (in the shelter of light))
  • Standard dilution stamp/samples: 24 mL (storage 4 °C or -20 °C)
  • HRP conjugate diluent: 10 mL (storage 4 °C or -20 °C)
  • Wash pad (25×): 24 mL (storage 4°C or -20°C)
  • TMB substrate solution: 12 mL (storage 4 °C or -20 °C (in the shelter of light))
  • Stopping solution: 7 mL (storage 4 °C or -20 °C)
  • Plate adhesive films: 2 pieces (storage 4 °C or -20 °C)
  • Instructions for use: 1 (storage of room temperature)

Documentation

Manufacturer SDS (ELK Biotechnology). Free download after signing in or creating an account.
Customers who have ordered: the SDS is attached automatically to the order confirmation.
A question or a specific configuration?
Contact us for a quote.