| Marque | Thermo Scientific |
| Réf. B-LABO | BL-EQ-173792 |
| Réf. fabricant | 30245785 |
| Adaptateur secteur | Universal 100-240 V 50/60 Hz power supply with wall plugs for US/Japan, Euro, UK/Singapore, Australia/New Zealand and China outlets |
|---|
| Sorties | USB flash drive, computer or printer (STARA-106) |
|---|
| Précision (baromètre) | ±6.0 mmHg; ±0.008 bar |
|---|
| Précision (DO) | ±2% saturation; ±0.02 mg/L or ppm |
|---|
| Précision (ions spécifiques) | ±0.2 mV or ±0.05% of the reading, whichever is greater |
|---|
| Précision (mV) | ±0.2 mV or ±0.05% of the reading, whichever is greater |
|---|
| Précision (rH) | ±0.2 mV or ±0.05% of the reading, whichever is greater |
|---|
| Précision (pH) | ±0.002 pH |
|---|
| Précision (OD par mesure polarographique) | ±2% saturation; ±0.02 mg/L or ppm |
|---|
| Précision (RDO) | ±2% saturation; ±0.02 mg/L or ppm |
|---|
| Fonctions d’ionométrie avancées | Single known addition, single known subtraction, double known addition, double known subtraction, analyte addition and analyte subtraction |
|---|
| Sorties d’alarme | Calibration reminder alarm, high and low limit alarm per parameter |
|---|
| Option de rétro-éclairage | Yes |
|---|
| Correction de la pression barométrique | Yes, automatic with built-in barometer or manual |
|---|
| Étalonnage | pH, ORP, Ion Concentration, DO, Barometric Pressure, Temperature |
|---|
| Modes d’étalonnage | Water-saturated air (100% and 0% saturation), air-saturated water (100% and 0% saturation) or custom concentration value (Winkler) |
|---|
| Points d’étalonnage | 1 to 5 (pH), 1 (ORP), 1 to 5 (ion), 1 to 2 (DO), 1 (barometer), 1 to 2 (temperature) |
|---|
| Points d’étalonnage (ionométrie) | Up to 5 |
|---|
| Points d’étalonnage (pH) | 1 to 5 points |
|---|
| Certifications/Conformité | CE, TUV 3-1, FCC Class A |
|---|
| Points de données | 10,000 with Date and Time Stamp |
|---|
| Canaux | 2 |
|---|
| Type d’affichage | 7-in. High-resolution Color Graphic LCD |
|---|
| Indice IP | IP54 |
|---|
| Measurement Modes | pH, ORP, mV, Ion Concentration (ISE), Dissolved Oxygen (mg/L), Dissolved Oxygen (% Saturation), Temperature, Barometric Pressure |
|---|
| Percent Saturation Range Polarographic DO | 0.0 to 600.0% saturation |
|---|
| Gamme (mV) | -2000 to 2000 mV |
|---|
| Gamme (pH) | -2 to 20 |
|---|
| Résolution (ionométrie) | 0.0001 minimum, 1 to 5 digits |
|---|
| Résolution (pH) | 0.1, 0.01 or 0.001 pH |
|---|
| Plage de températures (métrique) | -30.0°C to +130.0°C |
|---|
| À utiliser avec (application) | Most common pH samples as well as pH samples containing tris buffers, proteins, sulfides, halides or other components that can interfere with silver-chloride; wide variety of dissolved oxygen samples using optical DO electrode |
|---|
| Comprend | Meter with stand (PSTAR2160); ROSS Ultra Triode refillable epoxy-body pH/ATC electrode (8157BNUMD); optical DO/ATC electrode with calibration sleeve, guard and optical cap (087010MD); pH 4, 7, 10 buffers and rinse solution pouches; pH electrode storage solution and cleaning solution; pH electrode storage sleeve and base |
|---|
| Entrées | Channel 1: BNC for pH, ORP or ISE, 8-pin MiniDIN for ATC, pin-tip for reference; Channel 2: 9-pin MiniDIN for dissolved oxygen with ATC |
|---|
| Point isopotentiel | 7.000 pH; 1.000 ion concentration |
|---|
| Clavier | Capacitive touchscreen user interface |
|---|
| Langues | Chinese, English, French, German, Italian, Japanese, Korean, Portuguese, Spanish |
|---|
| Type de fonction d’enregistrement | Automatic with Auto-Read and Timed Read Modes, Manual with Continuous Read Mode |
|---|
| Mémoire | Non-volatile |
|---|
| Méthodes | 10 per channel (20 total) |
|---|
| Modèle | PD216 |
|---|
| Percent Saturation Relative Accuracy (RDO) | ±2% saturation |
|---|
| Percent Saturation Resolution (Polarographic DO) | 0.1, 1% saturation |
|---|
| Plage de saturation en pourcentage (RDO) | 0.0 to 600.0% saturation |
|---|
| Percent Saturation Relative Accuracy (Polarographic DO) | ±2% saturation |
|---|
| Résolution du coefficient de saturation (RDO) | 0.1, 1% saturation |
|---|
| Gamme de pression | 400.0 to 850.0 mmHg; 0.533 to 1.133 bar |
|---|
| Type de sonde | Polarographic or optical DO electrode |
|---|
| Type de produit | pH/Ion/DO Bench Meter pH and Optical DO Kit |
|---|
| Gamme (OD) | 0.0 to 600.0% saturation; 0.01 to 90.0 mg/L or ppm |
|---|
| Gamme (ion) | 0.0001 to 19999 |
|---|
| Gamme (ions spécifiques) | 0.0001 to 19999 |
|---|
| Gamme (rH) | -2000 to 2000 mV |
|---|
| Gamme (OD par mesure polarographique) | 0.0 to 600.0% saturation; 0.01 to 90.0 mg/L or ppm |
|---|
| Gamme (RDO) | 0.0 to 600.0% saturation; 0.01 to 90.0 mg/L or ppm |
|---|
| Gamme (mV relative) | -2000 to 2000 mV |
|---|
| Résolution (OD) | 0.1, 1% saturation; 0.01, 0.1 mg/L or ppm |
|---|
| Résolution (ion) | 0.0001 minimum, 1 to 5 digits |
|---|
| Résolution (mV) | 0.1 mV |
|---|
| Résolution (rH) | 0.1 mV |
|---|
| Résolution (OD par mesure polarographique) | 0.1, 1% saturation; 0.01, 0.1 mg/L or ppm |
|---|
| Résolution (RDO) | 0.1, 1% saturation; 0.01, 0.1 mg/L or ppm |
|---|
| Facteur de salinité | 0.0 to 50.0 ppt |
|---|
| Précision de la température | ±0.3°C, ±0.5°F |
|---|
| Étalonnage de la sonde de température | 1 or 2 point temperature offset calibration |
|---|
| Plage de températures (impériale) | -22.0°F to +266.0°F |
|---|
| Résolution de la température | 0.1°C, 0.1°F |
|---|
| Sélection de la température | Automatic with ATC temperature sensor, ambient or manual |
|---|
| Garantie | 2 year meter warranty |
|---|
| Dimensions (L x l x H) | 218.9 x 153.2 x 91.33 mm; 8.62 x 6.03 x 3.60 in. |
|---|
| Poids (impériale) | 2.20 lbs. |
|---|
| Poids (métrique) | 1.00 kg |